Mistry, Surajit and Kundu, Debabrata and Datta, Someswar and Basu, Debabrata and Soundrapandian, C (2011) Indigenous hydroxyapatite coated and bioactive glass coated titanium dental implant system - Fabrication and application in humans. Journal of Indian Society of Periodontology, 15 (3). pp. 215-220. ISSN 0972-124X

[img]
Preview
PDF - Published Version
Download (631Kb) | Preview

Abstract

Background: The use of different bioactive materials as coating on dental implant to restore tooth function is a growing trend in modern Dentistry. In the present study, hydroxyapatite and the bioactive glass-coated implants were evaluated for their behavior in osseous tissue following implantation in 14 patients. Materials and Methods: Bioactive glass and hydroxyapatite formulated and prepared for coating on Ti-6Al-4V alloy. Hydroxyapatite coating was applied on the implant surface by air plasma spray technique and bioactive glass coating was applied by vitreous enameling technique. Their outcome was assessed after 6 months in vivo study in human. Results: Hydroxyapatite and bioactive glass coating materials were nontoxic and biocompatible. Uneventful healing was observed with both types of implants. Conclusion: The results showed bioactive glass is a good alternative coating material for dental implant.

Item Type: Article
Uncontrolled Keywords: Bioactive glass, bioactive materials, biocompatible, hydroxyapatite, osseous tissue